Particle Counter CL 1070

    Deskripsi produk:

    Particle Counter Cl-1070
    Channel Sizes           : 0.5 µm, 1µm, 3µm, 5µm, 10µm, and 25µm
    Flow Rate                  : 100 LPM
    Battery / Electrical  : 3.0hours
    Size Calibration        : < ±10% with internal pulse height analyzer (PHA) in compliance with ISO 21501-4
    Size Resolution         : ±4% to ±10% typical
    Sample Volume Uncertainty : Sample Volume is user-adjustable in cubic meters, cubic feet, or liters.
    Data Interface           : RS-232, Ethernet, and USB.
    Sanitation                  : Compatible with all standard cleanroom sanitation and sterilization procedures,   including Vaporized Hydrogen Peroxide (VHP).
    Display                       : Large touchscreen display and large lettering. Easy to use interface. VERY quick response between screens—No frustrating delays.
    Alarms                        : Triggers at (i) Counts (User defined “Alert” and “Action” levels), (ii) Laser Status, and (iii) Flow Rate Deviation. 360 degree, user-configurable status and alarms (enable/disable, 5-colors, patterns, etc.)
    Programmable          : Five Security Levels, 100 Programs, 100 User ID’s plus administrator with password security, 100 location IDs, and up to 10,000,000 samples. All models uniquely come standard with Unit-to-Unit cloning.
    Environmental          : Temperature: 0-36oC (32-97oF). Humidity: 10-90% relative humidity, non-condensing
    Included Accessories : Internal thermal printer, internal battery charger, battery, stainless steel or light blocker isokinetic probe, sample tubing, zero count filter, two rolls of thermal paper, IQ/OQ documentation, manual, external exhaust fitting, and USB flash drive.