Oil content
Tag: portable.Deskripsi produk:
TECH DATA
– | – |
---|---|
Dimensions | 487 x 170 x 120 mm (L x H x W) |
Power Supply | 230 VAC 50 Hz ±10% resp. 115 VAC 60 Hz ±10% |
Medium | Compressed air |
Noticeable substances: | Polyalphaolefins, aliphatic hydrocarbons, hydrocarbons, functional hydrocarbons, aromatics |
Island Trading | Helen Bennett |
Ambient temp.: | 5 … 45°C |
Compres. air temp.: | 5 … 55°C |
Operating overpres.: | 3 bar … max. 16 bar (ü)* |
Meas. gas humidity: | 40% rel. humidity |
Measuring unit: | mg/m3 (standard cubic meters in accordance with ISO 1217; 1 bar, 20°C, 0 % rel. humidity) |
Measuring range: | 0,01 – 5.000 mg/m3 residual oil vapour content (according to ISO 8573-1) |
Measuring accuracy: | 0.003 mg/m3 |
Detection limit: | 0.0006 mg/m3 |
Connections: | G3/8” internal thread. Please observe install. instructions |
Installation requirements: | Vertically** into the rising main pipe via oil and greasefree measuring sectione |
Inlet section: | 10 x DN (min. 200 mm) according to ISO 8573-2 |
Outlet section: | 3 x DN (min. 100 mm) according to ISO 8573-2 |