    TECH DATA

    Dimensions 487 x 170 x 120 mm (L x H x W)
    Power Supply 230 VAC 50 Hz ±10% resp. 115 VAC 60 Hz ±10%
    Medium Compressed air
    Noticeable substances: Polyalphaolefins, aliphatic hydrocarbons, hydrocarbons, functional hydrocarbons, aromatics
    Island Trading Helen Bennett
    Ambient temp.: 5 … 45°C
    Compres. air temp.: 5 … 55°C
    Operating overpres.: 3 bar … max. 16 bar (ü)*
    Meas. gas humidity: 40% rel. humidity
    Measuring unit: mg/m3 (standard cubic meters in accordance with ISO 1217; 1 bar, 20°C, 0 % rel. humidity)
     Measuring range:   0,01 – 5.000 mg/m3 residual oil vapour content (according to ISO 8573-1)
     Measuring accuracy:  0.003 mg/m3
     Detection limit:  0.0006 mg/m3
     Connections:  G3/8” internal thread. Please observe install. instructions
     Installation requirements:  Vertically** into the rising main pipe via oil and greasefree measuring sectione
     Inlet section:  10 x DN (min. 200 mm) according to ISO 8573-2
    Outlet section: 3 x DN (min. 100 mm) according to ISO 8573-2